Twelve-channel ECG Machine MT-9112
Features
-
Digital signal processor for effective inhibition of baseline drift, AC interference filter, EMG interference filter and the heart rate, to guarantee the authenticity and dependability.
Specification
Lead: Standard 12 Leads
Acquisition Mode: Simultaneously 12 leads acquisition
Record Mode: manual mode, auto mode1, auto mode2, auto mode3
Filter: AC filter, EMG filter
CMRR: >100dB (with filter)
Input Circuit: floating input circuit, with protection circuit against defibrillation
Input loop Current: ≤0.1μA
Input Impedance: ≥2.5MΩ
Patient Current Leakage: ?10μA
Calibrating Voltage: 1mV?2?
Skin Voltage Tolerance: ?500mV
Time Constant: ≥3.2s
Frequency Response: 0.05?150Hz(-3dB)
Noise Level: ?15μVp-p
Sensitivity Threshold: ≤20μV
Inter-channel Interference: ≤0.5mm
Paper Speed: 12.5?25?50mm/s??3??
Sensitivity: AUTO, 2.5, 5, 10, 20mm/mV
Recording System: Thermal-array System
Recording Paper Specifications: 216mmx30m, Roll Paper
LCD: 5.7", 320?240 LCD, Display waveform and working status
Safety Standard: GB9706.1-1995, IEC class I, type CF
Power Supply: 110?230V(?10%), 50/60Hz?1Hz; 45VA
DC: 14.4V (2000mAh) rechargeable battery
Updated by: 10/21/16