Basic Info.
Model NO.: CMT-8000G
Type: Patient/Fetal/Maternal Monitor
Classification: Physiological Functions of Diagnosis and Monitoring Equipment
Export Markets: South America, Southeast Asia, Africa, Oceania, Mid East, Eastern Asia
Product Description
Standard configuration
15 inch TFT; Resolution: 1024*768
ECG, SpO2, PR, NIBP, RESP, TEMP
Function: Standard configuration (Include ECG, RESP, TEMP, PR, SpO2, NIBP), Built-in Lead-acid battery, one more battery, thermal recorder, touch screen, IBP, ETCO2, fetal monitor and Anesthetic depth can be optional. High light LED double tubes design, adopt Japanese good air tightness and low noise air pump, and the cable are both medical environmental materials, complete function, 15 inch display and unique Anesthetic depth monitoring make it can be widely used on the surgery and diagnosis.
Characteristics
1. Patent: Unique two modulator tube design, the display can work even one of that can not be used, this insure the display long-life usage
2. Safe: Adopt PU/TPU environmental cable and ensure the users' safety
3. Display: Import the Japanese SHARP TFT LCD display and ensure the high-light effect and no light-spot and waviness
4. Battery: One more battery can be optional, battery jar has good resilience and tightness.
5. Gold-plating probe terminal: Solid and durable, especially ensure the users' safety
6. High quality components and parts: Import Japanese air pump, good air tightness and low noise, American SMI pressure sensors make the measurement result more accurate.
7. Working environment: Wide range of temperature and humidity
8. Operate: Single key operating main function make the monitor easy
9. Optional Parameters
Thermal Recorder, Lead-acid Battery, IBP, ETCO2, Monitor depth of anesthesia, Anesthesia, gas monitoring, Oxygen monitoring
#ECG
Leads Mode; 5 leads; I\II\III\Avr\Avl\Avf\V
Gain(mm/mV): 1/4, 1/2, 1, 2, 4
Scanning speed; 12.5mm/s\25mm/s\50mm/s
Frequency response: Diagnosis 0.05-130Hz
Monitoring 0.5 - 40Hz
Operation 1.0- 20Hz
Heart rate range: 15-350 BPM
Heart rate accuracy: ± 1%
ST segment analysis: -0.8-? 0.8Mv
ST segment accuracy: ± 0.02Mv
#NIBP
Measure and Alarm Range
Adult Mode
SYS 40 ~ 280 mmHg
DIA 10 ~ 220 mmHg
MEAN 20 ~ 240 mmHg
Pediatric Mode
SYS 40 ~ 220 mmHg
DIA 10 ~ 160 mmHg
MEAN 20 ~ 170 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Static pressure range: 0~300mmHg
Static pressure accuracy: ± 3mmHg or 2%
Resolution: 1mmHg
Protection mode; Dual over voltage protection
#SpO2
Measure range: 1-100%
Accuracy: ± 2% (70-100%)
Resolution: 1%
#Pulse rate
Measure range: 20-300 BPM
Accuracy: ± 2 BPM
Resolution: 1 BPM
#Temperature
Channel; 2
Measure range: 0-50 centigrade
Accuracy: ± 0.1 centigrade
Resolution: 0.1 centigrade
#Respiration
Measure range; Adult 0~120BrPM
Pediatric/ Neonatal 0~150BrPM
Accuracy: ± 2%
Resolution: 1 BPM
#IBP(optional)
Accuracy (no sensor); ± 2%or1mmHg
Band width: Normal mode: DC~40Hz
Smooth mode: DC~12.5Hz
#ETCO2
Measure range: 0% - 13%
Resolution: 1 mmHg
Accuracy: ± 2 mmHg @ < 5.0% CO2 (at ATPS)
Respiration: 3 - 150 bpm
#Depth of anesthesia (optional)
EEG sensitivity ± 400µ V
Noise < 2µ Vp-p, < 0.4µ V RMS, 1-250Hz
CMRR >100dB
Input impedance >50Mohm
Sample rate 2000 samples/sec. (14 bits equivalent)
CSI and update 0-100. Filter 6-42Hz, 1 sec. Update
EMG 0-100logarithmic. Filter75-85Hz, 1sec. Update
BS% 0-100%. Filter 2-42 Hz, 1 sec. Update
#Option
Lead-acid battery (extermal plug)
Recorder (built-in)
IBP
ETCO2
Monitor depth of anesthesia
Anesthesia gas monitoring
Oxygen monitoring